The BaroTROLL® is a low-cost, self-contained logger that measures barometric pressure. It contains highly precise barometric and temperature sensors, a data logger, and internal power in a 18.3mm (0.72 in) diameter housing. Our low-owered system practically eliminates the need to change batteries and guarantees long-term operation. Alternatively, the unit can be powered with external power. System integration is a snap with built-in RS-485/232 Modbus, SDI-12, and 4-20mA communications.

The BaroTROLL® is typically used in conjunction with one or more of the non-vented Level TROLL® instruments to compensate for changes in water level due to barometric fluctuations. Readings from the BaroTROLL® can be automatically subtracted from data collected by a Level TROLL®. Due to its small size, it can be installed at the surface or inside the well casing--even in 25mm (1 in) wells! Built entirely from rugged titanium and welded to eliminate o-ring seals, you can have total confidence in harsh environments.

Applications:

  • Barometric pressure monitoring
  • Temperature monitoring
  • Atmospheric research
AWDS Logo

 

Request info


Features and Benefits:

  • Barometric pressure, temperature and data logging in a 18.3mm (0.72 in) OD – fits in a 25mm (1 in.) well.
  • Ultra-rugged, all-titanium design – perfect for corrosive sites.
  • TIG-welded seals – prevents flooding of instrument.
  • Ultra-low power system guarantees 5 years or 2 million readings – forget about changing batteries or running out of power.
  • Uses external battery or line power – never run out of power.
  • Sealed Twist-Lock hanger – 100% reliability.
  • Easy-to-use Win-Situ® software – Get the data you need fast! Large, sizable display that's easy to see from a distance.
  • Extremely accurate sensors – record minute changes in barometric pressure and temperature.
  • Data saved by sites not files – Collect and save information by site automatically. No need to transfer data later.
  • Easy-to-read memory and battery gauges – know exactly how much power is left.
  • Modbus, SDI-12, and 4-20mA – provides for easy connections to existing and online systems.
  • Specify start and stop times for logging – collect the data you need up to the date you want and no more.
  • User-friendly Wizards guide every step – reduces complicated steps and the need for tedious manuals.
  • -20 to 80C temperature range – use for research and applications with higher temperatures.
  • 100,000 data points of memory (1 MB) – reduces the interval between data collection.
  • CE and FCC approved – prevents susceptibility to outside emission interference and transient electrical surges.
  • Ability to "undelete" files from the unit – retrieves accidentally deleted data.

Specifications:

General  
Operating temperature range -20 to 80C
Dimensions 18.3 mm (0.72 in.) OD, 21.6cm (8.5 in.) long
Weight (includes backshell) 0.197 kg (0.43 lb.)
Output options Modbus (RS-485/232), SDI-12, 4-20mA
Housing material Titanium
Nose cone material Black Delrin®
Power
Internal battery 3.6V Lithium
Typical life* 5 yrs or 2 million readings
External power 8-36VDC
*Whichever comes first. See battery guide for details.
Logging
Memory / data points** 1 MB / 100,000+
Logging Rate User selectable from 1 min -- 49 days
Modbus 2 per sec
SDI-12 2 per sec
4-20 mA Update Rate 2 per sec
Measurement Types Linear
**data record = time and date plus one single parameter logged
Barometric Sensor Silicon strain gauge
Type Non-vented
Material Titanium
Accuracy***, FS +/-0.05% (at 15C)
FS = Full Scale +/-0.1% (-5 to 50C)
+/-0.25% (-20 to -5 & 50 to 80C)
*** Accuracy with 4-20mA output option: +/-0.25% typical
Resolution 0.005% FS or better
Range Non-vented: 0 to 1.14 bar (0 to 16.5 psi)
Units of Measure - Pressure psi, KPa, bar, mbar, mmHg, cmH2O, inH2O
Temperature Sensor
Material Silicon
Accuracy +/-0.1C
Resolution 0.01C or better
Range -20 to 80C (-4 to 176F)